Frequency Compensation
You are suggested to compensate the probe to match with the oscilloscope. The probes can
be adjusted for both low-frequency compensation and high-frequency compensation. Low-
frequency compensation is required at its first connection with the oscilloscope and high-
frequency compensation can be performed regularly.
Low-frequency compensation Cable Length
1 Connect the probe to the calibration signal output terminal and signal input terminal at the
front panel of the oscilloscope.
2 Set the probe attenuation ratio to 10X and press Auto at the front panel of the oscilloscope.
A 1 kHz square waveform is displayed on the screen.
3 Use the adjustment tool to adju