首页|嵌入式系统|显示技术|模拟IC/电源|元件与制造|其他IC/制程|消费类电子|无线/通信|汽车电子|工业控制|医疗电子|测试测量
首页 > 分享下载 > 测试测量 > LAB WM309 - Virtual Memories in WaveMaster

LAB WM309 - Virtual Memories in WaveMaster

资料介绍
Here is a way to make more than 4 memory channels
LeCroy Applications Brief No.LAB WM309

Virtual Memories in WaveMaster
Here is a way to make more than 4 memory channels.
Measuring skew between multiple
waveforms often requires the stor-
age of many waveforms for recall
and measurement. However, most
digital scopes only have two or
four waveform memories for trace
storage. This small number of
memory locations limits the num-
ber of traces recalled into the
scope simultaneously. For applica-
tions where measurements must be
made on more than 4 traces, (such
as skew measurements in memory
chips, or propagation delay in
fanout buffers), a more powerful
solution must be used.

The following example uses the
customization capability in the
WaveMaster to create 8 virtual Figure 1
LAB WM309 - Virtual Memories in WaveMaster