Abstract: This application note introduces how a continuous positive airway pressure (CPAP) ventilation system operates. The main subfunctions of CPAP respiratory ventilation include air-hose-environment sensing, compressor motor-drive feedback, motor-drive excitation, and a communication interface to a technician/doctor. These subfunctions are explained and a functional block diagram showing system components is also detailed.
Maxim > Design support > App notes > A/D and D/A Conversion/Sampling Circuits > APP 4685
Maxim > Design support > App notes > Amplifier and Comparator Circuits > APP 4685
Maxim > Design support > App notes > Sensors > APP 4685

Keywords: cpap, continuous positive airway pressure, block diagram, electrical components, respiratory ventilation

May 10, 2010
APPLICATION NOTE4685
How Continuous Positive Airway Pressure (CPAP) Respiratory
Ventilation Systems Function
By:John Gosson

Abstract:This application note introduces how a continuous positive airway pressure (CPAP) ventilation system operates. The main
subfunctions of CPAP respiratory ventilation include air-hose-environm
How Continuous Positive Airway Pressure (CPAP) Respiratory Ventilation Systems Function