资料介绍
Abstract: This tutorial discusses how the proper design of trim, adjustment, and calibration circuits can correct system tolerances, making medical equipment safer, more accurate, and more affordable. Calibration subjects addressed include compensating for component tolerances, using final-test calibration, improving reliability through power-on self-test and continuous/periodic calibration, enabling accurate automated adjustments, replacing mechanical trims with all-electronic equivalents, and leveraging precision voltage references for digital calibration. Maxim > Design support > App notes > A/D and D/A Conversion/Sampling Circuits > APP 4703
Maxim > Design support > App notes > Digital Potentiometers > APP 4703
Maxim > Design support > App notes > Voltage References > APP 4703
Keywords: medical instruments, calibration, trim, electronic, self test, compensation, digital pot, potentiameter, replace mechanical,
CDAC, CDPOT, calibration dac, calibration pot, digtial calibration
May 10, 2010
APPLICATION NOTE4703
Introduction to Electronic Calibration and Methods for Correcting
Manufacturing Tolerances in Medical Equipment Designs
By:Bill Laumeister, Strategic Applications Engineer
Abstract:This tutorial discusses how t