Abstract: This application note provides an introduction to ventilators and their basic functions. It discusses how ventilation rates are measured and provides the formula for determining if the tidal volume of gas exchange is sufficient for the patient. The system subfunctions addressed include air-oxygen mixture sensing, air-oxygen mixture control, inspiration control, the communication interface with the technician/doctor, and the alarm system.
Maxim > Design support > App notes > A/D and D/A Conversion/Sampling Circuits > APP 4692
Maxim > Design support > App notes > Amplifier and Comparator Circuits > APP 4692

Keywords: ventilation, ventilator, gas exchange, tidal volume, electrical components, medial equipment, block diagram

May 10, 2010
APPLICATION NOTE 4692
Introduction to Ventilators and Ventilation Rate Measurements
By: John Gossan

Abstract: This application note provides an introduction to ventilators and their basic functions. It discusses how ventilation rates are
measured and provides the formula for determining if the tidal volume of gas exchange is sufficient for the patient. The system
subfu